Desfibrilhador Defi Xpress

Defibrillator Defi Xpress

09DFB005
€4,800.00
Tax excluded

 

Defi Xpress Defibrillator

With voice alarm that can meet your clinical needs from Basic Life Support (BLS) to Advanced Cardiac Life Support (ACLS).

5 Available languages: English, Russian, Spanish, Chinese, Turkish

The Shredders are for patients who experience a sudden case of heartbeat and arrhythmias. In these cases

The shredders are an important tool for anyone in the vicinity of a victim who suffers from ventricular arrhythmia as well. arrhythmia makes the patient unconscious. the heart may beat too fast, too slow, or an irregular mixture of both. The expected result of arrhythmias is cardiac arrest or a stroke.

13 cm
6 cm
33 cm
5 kg
New

Screen

Screen type: High resolution screen

Screen size: 7 inches (17.7 cm) diagonal

Scan speed: 25 mm / sec

Information: Heart Rate. Lead / pads. Alarm On Off. SpO2 AED

Functions and commands selection of alarms and limits of supplied power.

5 languages available: English, Russian, Spanish, Chinese, Turkish

 

Defibrillator

Waveform: biphasic

Charging time: Less than 7 seconds with a new fully charged battery.

Power Display: Monitor display indicates selected and supplied power.

Charging Prompt Type: Voice and visual prompts.

Electrode Impedance Measurement Range: 0-250 ohms.

 

ECG Monitoring

Patient Connection: 3-lead ECG cable or 5-lead ECG pads.

Lead Selection: displayed on monitor pads I II III AVR AVL AVF V.

ECG Size: 0.25 0.5 1 1.5 2 4 cm / mV on the monitor.

Heart Rate: 20-300BPM.

Heart Rate Alarm: on / off displayed on the monitor user selectable.

Smart Alarms: beep / voice prompts indicate shock rate.

 

Battery

Type: 12V chargeable Ni-MH battery.

Operating time: For a new fully charged battery: 60 defibrillator discharges at maximum power or 3 hours minimum continuous ECG monitoring. Additional parameters will charter the operating time with different functions.

 

SpO2 Module (Options):

Measurement Range: 30 ~ 100% ± 2% between 80% ~ 90% others ± 5%.

Alarm Range: User defined high limit and low limit.

Alarm accuracy: ± 10% within the setting values.

Alarm Time Accuracy: less than 12 sec.

Pulse Oximeter 1-Meditech

Pulse Oximeter 2-Nellcor

 

Printer:

Paper: 50mm thermal.

Speed: 12.5 mm / sec 25 mm / sec 50 mm / sec. Delay of 6 seconds user selectable.

Printing Method: High resolution thermal print head

Printing modes: Manual or automatic user configurable.

On / Off control: Front panel.

Automatic function: 9 seconds recording initiated by alarm activation or defibrillator charge or defibrillator discharge.

 

AED Mode

AED Function: Automatically analyze and charge x3 with programmable automatic energy level selection screen prompts and voice commands.

Shockable rhythms: Ventricular fibrillation with amplitude> = 200uV ventricular tachycardia with rates> = 140bpm and QRS complex wave length> = 140ms.

Charge Control: Control on the front panel of the device press the key on the paddle.

Prompts: Voice and visual prompts.

 

Manual mode

Power Selection: Selectable in 2 5 7 10 20 30 50 70 100 150 200 300 360 joules.

Synchronized Mode: Synchronizes the pulse of the defibrillator with the R-Wave "SYNC" message displayed on the monitor.

The specification of the battery's electromagnetic compatibility of the physical isolation defibrillator and the performance of the rhythm recognition will be described in this section:

Data Acquisition and Analysis

The DEFI XPRESS is no longer automatically configured for a particular frequency response. There are two diagnostic frequency options: 0.05-150 Hz

Diagnostic frequency filter and filtered diagnosis from 0.05 to 40 Hz

Frequency filter. See chapters 7 on how to switch between these two options.

 

Defibrillator

Waveform details:

The tables below provide details of the waveforms provided by the DEFI XPRESS when connected to resistive loads of 25 50 and 100 Ohms and set to their maximum output. The waveforms are characterised by DEFI XPRESS values for peak current (Ip) for two-phase: duration of the first output phase (tphase1) and duration of the second output phase (tphase2). The values shown are within 10%.

Waveform: truncated exponential biphase

 

Output power accuracy:

+/- 10% or 1J (whichever is greater) at 50 ohms

+/- 15% or 1J (whichever is greater) between 25 and 100 ohms

 

Power selection:

External: 2 5 7 10 20 30 50 70 100 150 200 300 360 J.

 

Charging time:

<12 seconds @ 360 J (with a 12 volt Ni-MH battery - MEDITECH after 15 discharges).

<15 seconds @ 360 J (without 90% AC mains voltage from the battery)

Charge indicator: audible and graphic.

 

Output:

Pediatric adult paddle adapter and multi-purpose hands-free adapter available.

Synchronizer: Provides power within 60 ms.

Disarm: Front panel switch.

Monitor / Display

Input:

5-lead and 3-lead patient cable

Size: 7 inches (16.5 cm) diagonally non-fading

Type: colour (TFT)

Monitor resolution: 800 x 600 pixels.

Scan speed: 12.5 mm / s25 mm / s50 mm / s

Lead Selections: I II III AVR AVL AVF V.

Frequency Response: (User selectable).

Operating mode from 0.7 to 20 Hz

Monitoring mode 0.3 to 40 Hz

Analysis mode 0.05 to 150 Hz

AED mode from 0.5 to 30 Hz

Common mode rejection:> = 60dB

Heart Rate Meter: 20 to 300 BPM.

Heart Rate Alarms: User selectable.

Size: 0.125 0.25 0.5 1 2 4 cm / mv and automatic range.

ECG output: 1 V / mV.

Heart Rate Response to Irregular Rhythm: (AAMI EC13-1992 section 4.1.2.1.e.)

 

Alarms

Heart rate alarm Spo2:

Audible: 5 pulses of 800 tones with a PW of 150 msec per PRI of 225 msec and a repetition interval of 10 seconds.

Visual: The heart rate alarm causes the displayed heart rate to flash at 2 Hz. This display is located at the top of the display and is 0.4 "high and 0.28 to 0.840" wide, depending on the number of digits in the heart rate. Black is amber or whatever colour the display text is.

 

Battery failure alarm:

Audible: triplet tone of 3 pulses of 500 Hz with a PW of 200 msec per PRI of 310 msec. When the HR alarm is set or the defibrillator is at the lead failure tone, it repeats at a 20 second repetition interval.

Visual: Lead Fault condition causes a "LEAD FAULT" message to be displayed on the trace along with a dashed line across the width of the trace. The text is 0.18" high and is the colour of the display text. The dashed line is 5" Long display width. Signal is not modulated.

 

Physiological Alarms (Spo2):

Audible: View the heart rate alarm

Visual: Physiological alarms cause the displayed parameter to flash at 2 Hz. The colour is amber or equal to the coloured display text.

Duration mute: 90 seconds.

 

Printer:

Type: high resolution thermal matrix.

Annotation: Lead ECG time date ECG gain cardiac defibrillation and ACLS events summary of treatment.

Paper Width: 50 mm.

Paper Speed: 25 mm / sec 50 mm / sec 12 leads 50 mm / sec

Delay: 6 seconds

Frequency Response:

Automatically set to frequency response of the monitor.

 

Battery

High capacity type: Ni-MH 12V volt - MEDITECH

Up to 4 hours of ECG monitoring or 110 total energy discharge or 3 hours combined ECG Spo2 and monitoring

Proper battery care is required to maintain maximum available capacity.

Low Battery Indicator:

Low battery icon flashing on display and LED flashing on front panel.

Recharge rate: 80% in 7.5 hours. 100% in 9 hours.

 

General

Weight: 10 lbs (4.95 kg) (basic configuration).

Size: 13 x 12.5 x 5.3 inches (33 x 31.8 x 13.5 cm).

 

Operating:

Temperature *: 0 to 45 ° C

Humidity * (NC): 15 to 95% RH (30 to 90% with CO2 probe)

Vibration*: MIL-STD 810E

Shock/Drop*: MIL-STD 810E

Altitude: up to 4572 M (15,000 feet)

Transportation and Storage

Temperature: -30 to 70 ° C (-20 to 70 ° C with CO2 probe)

Humidity: 15 to 95% relative humidity (non-condensing)

Atmospheric Pressure: 860 to 1060 hPa

Shock / vibration: ISTA 1A

Enclosure protection *:

Solid foreign object: IEC 529 IP2X

Water: IEC 529 IPX4.

All tests performed per AAMI DF-2 defibrillation standard.

 

Operating Power:

12-volt Ni-MH - MEDITECH

SAED rhythm recognition performance

DEFI XPRESS SAED algorithm exceeds ANSI / AAMI requirements

Product added to wishlist

iqitcookielaw - module, put here your own cookie law text