I-PAD defibrillator
Easy to use with LED indicators, automatic self-test, truncated exponential biphasic waveform
Temperature: 0°-40°C
Humidity: 5%-90% non-condensing
Operating Mode: Semi-automatic
Waveform: biphasic e-cube (truncated exponential type); impedance compensated
Shock Energy Accuracy: Within ±3% (150J at 50Ω)
Charge Control: Automatic by Software default ;(Arrhythmia Detection and Charge Control System)
Charging time "Shock Advised": <14 seconds typically
Shock to Shock Cycle Time: <14 seconds typically including analysis
Battery Type: 12 Volt DC, 4.2 Ah, lithium manganese dioxide, long-life disposable primary cell
Capacity: Minimum 200 shocks or 4 hours operating time
Temperature: -40 °C to 60 °C
Defibrillator electrodes: Self-adhesive, disposable, non-polarized defibrillator electrodes
Adult Defibrillator Electrodes: Defibrillator electrodes for patients over 8 years of age or weighing more than 25 kg.
Surface Area:Adult: 110cm2 each
Cable Length: 1.5m