12 Inch TFT Patient Monitor - MD9012
12 inch TFT patient monitor
Automatic temperature drift handling of patented NIBP measurement technology
User-friendly and configurable user interface
Powerful data review: Maximum 720-hour graphical and tabular trends;
Up to 10000 NIBP log storage units
12" high resolution color TFT display
12 channel waveforms. 9 Parameters: ECG, Resp, NIBP, SpO2, TEMP,
Light weight, compact and portable
Built-in rechargeable lithium battery
Single human voice alarm
Advanced screen layouts: standard, large font, trend table, NIBP trend, full ECG lead, OxyCRG, IoC layout
Powerful storage: maximum 720 hours of all graphical and tabular trends
parameters, up to 10000 NIBP review, 200 alarm review, 2 Hours of all waveforms
System software can be updated over the network
Data and waveforms can be adjustable
Arrhythmia analysis and S-T segment analysis
Suitable for adult, pediatric and neonatal patients
Options
Invasive Blood Pressure (IBP)
Carbon dioxide (CO2) module
Thermal Printer
Touchscreen
ECG
Lead Mode 5 Leads (R, L, F, N, C ou RA, LA, LL, RL, V)
Lead selection I, II, III, avR, avL, avF, V,
2 ch waveform
Lead mode 3 Leads (R, L, F ou RA, LA, LL)
Lead Selection I, II, III
1 ch waveform
Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto
HR and Alarm
Range
Adult 15 ~ 300 bpm
Neo / Ped 15 ~ 350 bpm
Accuracy ± 1% or ± 1bpm, which is optimal
1 bpm Resolution
Sensitivity> 200 (uV)
Differential input impedance> 5 MΩ
CMRR
Monitor> 105 dB
Operation> 105 dB
Diagnosis> 85 dB
Electrode displacement potential ± 300mV
Leak Current <10 uA
Baseline recovery <3 S After Defi.
ECG Signal Range ± 8 m V (Vp-p)
Bandwidth
Surgery 1 ~ 15 Hz
0,5 ~ 35 Hz Monitor
Diagnosis 0.05 ~ 100 Hz
Calibration Signal 1 (mV), Accuracy: ± 5%.
ST Segment Monitoring Range
Measure and Alarm -2.0 ~ +2.0 mV
ARR detection
Type ASYSTOLE, VFIB / VTAC, PARES, BIGEMINY, TRIGEMINY, R ON T, VT> 2, PVC, TACHY, BRADY, MISSING BEATS, PNP, PNC
Alarm Available
Revision Available
RESPONSE
Method Impedance between R-F (RA-LL)
Input differential impedance > 2,5 MΩ
Measurement impedance range: 0.3~5.0Ω
Baseline impedance: 0 - 2.5 KΩ
Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate
Measurement Range and Alarm
Adult 0 ~ 120 rpm
Neo / Ped 0 ~ 150 rpm
1 rpm resolution
Precision ± 2 rpm
Apean Alarm 10 ~ 40 S
NIBP
Oscillometric Method
Manual, Auto, STAT mode
Measuring Range in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
Measurement Period in STAT Mode 5 Min
Pulse rate range 40 ~ 240 bpm
Alarm type SYS, DIA, MEAN
Measurement and alarm range
Adult Mode
SYS 40 ~ 270 mmHg
DIA 10 ~ 215 mmHg
AVERAGE 20 ~ 235 mmHg
Pediatric Mode
SYS 40 ~ 200 mmHg
DIA 10 ~ 150 mmHg
AVERAGE 20 ~ 165 mmHg
Neonatal mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
AVERAGE 20 ~ 110 mmHg
Resolution
1mmHg pressure
Precision
Pressure
Maximum mean error ± 5 mmHg
Maximum standard deviation ± 8mmHg
Overpressure Protection
Adult Mode 297 ± 3 mmHg
Pediatric Mode 240 ± 3 mmHg
Neonatal Mode 147 ± 3 mmHg
SpO2
Measuring range 0 ~ 100%
Alarm Range 0 ~ 100%
1% Resolution
Accuracy 70% ~ 100% ± 2%
0% ~ 69% not specified
Update interval of about 1 sec.
Alarm Delay 10 sec.
Pulse Rate
Measurement and Alarm Range 0 ~ 254bpm
1bpm Resolution
Precision ± 2bpm
TEMPERATURE
Channel 1
Measurement and Alarm Range 0 ~ 50 C
Resolution 0.1 ~ c
Precision ± 0.1 ~ C
Update interval of about 1 sec.
Average Time Constant <10 seg.
IBP
Label, PA, CVP, RAP, LAP, ICP, P1, P2
Alarm measurement and range
ARTE 0 ~ 300 mmHg
PA -6 ~ 120 mmHg
CVP / RAP / LAP / ICP -10 ~ 40 mmHg
P1 / P2 -10 ~ 300 mmHg
Press the sensor
Sensitivity 5 uV / V / mmHg
Impedance 300-3000Ω
1 mmHg Resolution
Accuracy of ± 2% or ± 1mmHg, which is great about 1 sec.
ECG
Lead Mode 5 Leads (R, L, F, N, C ou RA, LA, LL, RL, V)
Lead selection I, II, III, avR, avL, avF, V,
2 ch waveform
Lead mode 3 Leads (R, L, F ou RA, LA, LL)
Lead Selection I, II, III
1 ch waveform
Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto
HR and Alarm
Range
Adult 15 ~ 300 bpm
Neo / Ped 15 ~ 350 bpm
Accuracy ± 1% or ± 1bpm, which is optimal
1 bpm Resolution
Sensitivity> 200 (uV)
Differential input impedance> 5 MΩ
CMRR
Monitor> 105 dB
Operation> 105 dB
Diagnosis> 85 dB
Electrode displacement potential ± 300mV
Leak Current <10 uA
Baseline recovery <3 S After Defi.
ECG Signal Range ± 8 m V (Vp-p)
Bandwidth
Surgery 1 ~ 15 Hz
0,5 ~ 35 Hz Monitor
Diagnosis 0.05 ~ 100 Hz
Calibration Signal 1 (mV), Accuracy: ± 5%.
ST Segment Monitoring Range
Measure and Alarm -2.0 ~ +2.0 mV
ARR detection
Enter AS